Bookmark this page | Homepage | Casual look | Mobile version | 如有业务需要请点击QQ咨询如有业务需要请点击QQ咨询Skype ConsultingSkype Consulting
Member

PG Instruments Limited,.$

人民币

 
Products
  • No Category

Contact
  • Contact:Mr. April Cao
  • Phone:86-10-69910653
Search
 

riends links
  • No link
Your current location:Home » Product supply » T70 Optical Emission Split Beam Spectrophotometer, Photometric Range - 0.3 - 3.0Abs
T70 Optical Emission Split Beam Spectrophotometer, Photometric Range - 0.3 - 3.0Abs
Click image to view full size image
Product: Views:9T70 Optical Emission Split Beam Spectrophotometer, Photometric Range - 0.3 - 3.0Abs 
Unit price: Negotiable
MOQ:
Quantity:
Delivery date: Since the payment date Days delivery
Valid until: Long-term effective
Last updated: 2017-02-14 17:27
  Inquiry
Details

 

Optical Emission Split Beam Spectrophotometer T70, Photometric range - 0.3 - 3.0Abs


 

T70 is a high performance split beam spectrophotometer available with a fixed spectral bandwidth, which is innovative in terms of instrument application, mechanical and optical design, electronic control and software whilst retaining features that are well established and accepted through the industry.

T70 is able to carry out photometric measurement, spectrum scans, quantitative determination and DNA/Protein analysis. When interfaced to a PC using the UV-Win software, many more features are available including three dimensional spectrum, kinetic measurements, method and data storage, exportation of data in multiple formats and GLP administration features.

 


Features & Functions

  • High performance fixed 2nm spectral bandwidth
  • Wavelength accuracy +0.3nm
  • Supplied with a motorized 8 cell changer and pre-aligned Tungsten and Deuterium lamps
  • Holographic blazed grating 1200lines/mm
  • High degree of automation requiring minimal key depressions to start analysis
  • A number of optional accessories available which increase the flexibility of the analysis
  • Analysis for photometric measurement, spectrum scans, quantitative determination and DNA/Protein analysis
  • UV-Win software gives additional functionality including 3D spectrum analysis and compliance with GLP protocol.
  • Simple mechanical structure and modular electronics make routine maintenance very easy




Applications:
 

  • Education in universities and colleges
  • Environmental monitoring & control
  • Food & Beverage production and quality control
  • Agricultural monitoring
  • Geology & metallurgy
  • Gas & oil


 

Specifications:

 

Optical system

Split beam ratio

Scan speed

Selectable 

Wavelength range

190-1100nm

Wavelength accuracy

+0.3nm

Wavelength reproducibility

≤0.2nm

Spectral bandwidth

2nm

Photometric mode

Transmittance, Absorbance, Energy Concentration

Photometric range

-0.3-3.0Abs

Photometric accuracy

0.002A (0-0.5A), 0.004A (0.5-1A), 0.3%T (0-100%T)

Photometric reproducibility

00.001A(0-0.5A), 0.002A(0.5-1A), 0.15%T(0-100%T)

Photometric noise

0.001A(500nm) 30min warm-up

baseline flatness

0.002A(200-1000nm)

baseline stability

0.001A/h (500nm, 0Abs), 2hr warm-up

Stray light

≤0.12%T(220nm NaI, 340nm NaNO2)

Standard functionality

Photometric, Quantitative, Spectrum and DNA measurements

Cell holder

Automatic 8 cell changer

Detector 


Inquiry

(c)2015-2016 dahmw.org All Rights Reserved